| Model Number |
CF-EZ1500DL |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the colonovideoscope had a separate issue.During the initial evaluation of the device, a white residue was observed in the air water supply and the auxiliary water flow.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device, the white residue identified in the air water supply and the auxiliary water flow during the initial evaluation was confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the white residue in the air water supply and the auxiliary water flow could not be determined.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|