| Model Number |
SGC0801 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 that during preparation of the steerable guide catheter (sgc), when inserting the dilator into the hemostatic valve, a loss of fluid column occurred.Therefore, the sgc was not used and was replaced.There was no clinically significant delay in the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported leak during prep.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device returned for analysis, a cause for the reported leak during prep could not be conclusively determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|