The patient reported that after connecting a discharged controller to a charger, the device was left charging for approximately 35 minutes before an alarm was heard.Upon inspection, the charging cable had reportedly overheated and fused to the controller's charging port, leaving the controller unable to charge and displaying a battery level of 1%.The patient stated that a non-insulet usb charging cable was being used because the original insulet-supplied charging cable was lost.Medical attention was not required.The patient will be issued a new device.
|
|
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed and the product lot met all acceptance criteria.Cloud - locked down/smartphone,
data not available.Cloud - omnipod 5 software app version,
data not available.Cloud - smartphone operating system,
data not available.Cloud - smartphone hardware,
data not available.Cloud - cgm sensor type,
data not available.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|