• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Insulet Corporation Omnipod 5 Controller; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod 5 Controller; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number PT-000409
Medical Device Problem Codes Thermal Decomposition of Device (1071); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
The patient reported that after connecting a discharged controller to a charger, the device was left charging for approximately 35 minutes before an alarm was heard.Upon inspection, the charging cable had reportedly overheated and fused to the controller's charging port, leaving the controller unable to charge and displaying a battery level of 1%.The patient stated that a non-insulet usb charging cable was being used because the original insulet-supplied charging cable was lost.Medical attention was not required.The patient will be issued a new device.
 
Additional Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed and the product lot met all acceptance criteria.Cloud - locked down/smartphone, data not available.Cloud - omnipod 5 software app version, data not available.Cloud - smartphone operating system, data not available.Cloud - smartphone hardware, data not available.Cloud - cgm sensor type, data not available.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod 5 Controller
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50436859
Report Number3014585508-2026-60653
Device Sequence Number9167948
Product Code QFG
UDI-Device Identifier10385083000022
UDI-Public(01)10385083000022(11)230106(10)H000496
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203768
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPT-000409
Device Catalogue NumberPDM-H001-G-XX
Device Lot NumberH000496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/01/2026
Initial Report FDA Received Date08/25/2026
Date Device Manufactured01/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-