| Catalog Number |
00880100100 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, outside of surgery, a device issue was identified during a recall in which the device could cut beyond the intended depth due to a misaligned thickness control bar.During device evaluation, it was discovered that the rpms were unable to be taken due to the reciprocating arm was jumping erratically and would not settle for a reading.There was no patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined that the unit had product recall.Rpms were unable to be taken due to the reciprocating arm was jumping erratically and would not settle for a reading.The control bar was loose and not in position.The motor, ball bearings, and neck bearing were all replaced.The spring seal was stretched to raise rpms.New vespels were installed.The control bar was adjusted to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|