| Catalog Number |
364990 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Report 5 of 7.It was reported that when using the bd vacutainer® urinalysis transfer straw kit, an unspecified number of units broke while in the centrifuge, requiring recollection.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 17 samples and 5 photographs from the customer in support of this complaint for catalog number 364990, lot number 6048195.Evaluation of the photographs confirmed the customer's reported failure mode of damaged/broken tubes, specifically a tube with a damaged bottom.The 17 returned samples were visually inspected, and no damage or cracks were observed on the tube bottoms.The device history records were reviewed with no issues being identified.There were no related quality notifications.All processes and final inspections comply with specification requirements.This complaint has been confirmed for the reported failure mode of damaged/broken tubes.However, no definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|