During a rescue attempt involving a patient who had been removed from the water, the customer attempted to use the aed but experienced difficulty getting the defibrillation pads to adhere to the patient's wet, hairy chest.The fire department subsequently arrived, assumed responsibility for the rescue, and used its own equipment.The customer stated that the aed did not malfunction and clarified that the purpose of the inquiry was to replace the pads used during the rescue and potentially purchase additional spare sets.No information regarding the patient's condition or outcome was provided.
|
|
This event is being reported as a malfunction because the reported inability of the defibrillation pads to adhere adequately, if it were to recur, could delay or prevent the delivery of defibrillation therapy and could cause or contribute to a death or serious injury.The customer had disposed of the defibrillation pads used during the rescue attempt and did not provide the pad lot number or expiration date.The customer attributed the reported difficulty with pad adhesion to the patient's wet skin and excess chest hair.The ddu-120 series operating guide instructs users to remove clothing from the patient's chest and, if necessary, dry the chest and remove excess chest hair.The guide also warns that contact between the patient's body and conductive fluids"[?]"including water, gel, blood, or saline"[?]"or metal objects may provide unwanted pathways for defibrillating current.Because the pads were not returned and the lot number and expiration date were unavailable, the reported issue could not be reproduced or evaluated.Although the conditions reported by the customer may have contributed to the difficulty with pad adhesion, a definitive root cause could not be confirmed.
|