It was reported that approximately six months after implantation of an intraocular lens (iol) in the patient's dominant right eye (od), the lens was exchanged with a different model due to persistent halos experienced during both daytime and nighttime conditions.Prior to the lens exchange, the surgeon attempted to manage the symptoms by treating the patient's dry eye; however, the halos remained unresolved.In the surgeon's opinion, the most likely cause of the event was the patient's visual system/brain inability to tolerate halo phenomena associated with multifocal iols, noting that such symptoms can occur with all multifocal lenses.Additional information was requested but not received.
|
|
Although requested, the device was not returned for evaluation, and additional information was not received.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of the event could not be conclusively determined, however patient related factors may have caused or contributed to this event.
|