• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Stryker GmbH Gamma4 - Intermediate Targeting Sleeve; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Stryker GmbH Gamma4 - Intermediate Targeting Sleeve; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 14202100
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "on the (b)(6)2026 a g4 intermediate nail has been used at (b)(6) hospital.This case was covered by (b)(6).The distal locking screw through the gamma4 targeted jig missed the nail.This was believed to be corrected at the time by re-attaching the jig and checking on xray.However, at a follow up xray the screw has appeared to miss posteriorly compared to it previously missing anteriorly.This resulted in the patient coming back to theatre on the (b)(6)2026 to have another distal screw inserted through the nail.The jig was re-attached and again when drilling through the intermediate jig the drill was hitting the nail and not aligning with the locking screw hole.The surgeon mateo managed to get a screw through the oblique hole of the intermediate nail and tried again through the static locking.The issue was raised that he believes there may be some issues with the jig itself or the nail screw holes not aligning with the jig correctly if 1 screw is able to pass through but not the other.".
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Gamma4 - Intermediate Targeting Sleeve
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
Stryker GmbH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
Stryker Trauma GmbH
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key50440232
Report Number0008031020-2026-01347
Device Sequence Number15667233
Product Code HSB
UDI-Device Identifier07613327591477
UDI-Public07613327591477
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K230512
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number14202100
Device Lot NumberKP505180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-