• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT265
Device Problems Crack (1135); No Pressure (2994)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date no (b)(4) infant evaqua 2 breathing circuits have been returned to fisher & paykel healthcare (fph) in (b)(4).An investigation was conducted based on the photographs provided by an fph field representative.Visual inspection of these photographs revealed that the proximal connector collar was cracked on one expiratory limb.As no devices have been received for evaluation, a full investigation could not be completed.Therefore the cause of the observed damage could not be determined.Based on the nature of the observed cracking and previous investigations into this type of failure, this type of cracking can occur when the connectors come into contact with a cleaning chemical, resulting in environmental stress cracking.All rt265 infant dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.Our user instructions that accompany the rt265 state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.The user instructions also state the following: do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.
 
Event Description
A hospital in (b)(6) reported that the collar on the expiratory limb of the rt265 infant dual-heated evaqua2 breathing circuit was found cracked during patient setup which caused a loss in ventilatory pressure.This was found prior to patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key5044116
MDR Text Key25298277
Report Number9611451-2015-00388
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 08/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT265
Device Catalogue NumberRT265
Was Device Available for Evaluation? No
Date Manufacturer Received08/07/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-