• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Stryker-Endoscopy Lakeland 5300; Laparoscope, general & plastic surgery, reprocessed

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Stryker-Endoscopy Lakeland 5300; Laparoscope, general & plastic surgery, reprocessed Back to Search Results
Model Number D12LT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trocar broke while using on patient.The complainant is not aware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, only the cannula was returned for investigation.The cannula housing was separated in half at the adhesive line, with adhesive present.A review of the dhr supports that the device met all inspections and test criteria prior to release from stryker.The reported event could be attributed to: - excessive force applied.- contact with hard object or other improper handling.- insufficient structural integrity.- shipping/handling damage or extreme conditions.The instructions for use (ifu) state: - prior to use, read and follow the instructions of this insert as well as those of the instruments to be used during the procedure.Damage to the instrument can lead to patient injuries.Always inspect instrument carefully for overall integrity before use.- keep the trocar straight relative to the cannula when inserting or removing.If the trocar is at an angle relative to the cannula, it can damage the cannula and result in desufflation.- do not use excessive force.- verify compatibility of all instruments before use to avoid complications during surgery.- become familiar with specific model of trocar and cannula prior to employing it in a surgical procedure to avoid damage to patient, to operator or to instrument.- careful handling of instruments is necessary to avoid damage or breakage.- care should be taken when removing instruments not to prematurely dislodge the cannula.The reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
Laparoscope, general & plastic surgery, reprocessed
Manufacturer (Section D)
Stryker-Endoscopy Lakeland 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
Stryker-Endoscopy Lakeland 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
arsalan rizvi
1810 w. drake drive
tempe, AZ 85283
+1(888)888-3433
MDR Report Key50444280
Report Number0001056128-2026-00063
Device Sequence Number9780099
Product Code NLM
UDI-Device Identifier00885825015596
UDI-Public(01)00885825015596(17)290109(10)17417496
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K201511
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD12LT
Device Catalogue NumberD12LT
Device Lot Number17417496
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/21/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/28/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-