| Catalog Number |
0240200100 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was overheating.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: overheating warning.The failure identified in the investigation is consistent with the complaint record.The probable root cause is due to a large lint ball preventing the base intake fan blade from functioning; possibly leading to the overheating of the console.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|