| Model Number |
380614-15 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the customer called a field service engineer (fse) and reported that the system had an error 25596 on the armnet1 remote arm controller 1 (rac1).The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: error could not be resolved, arm1 disabled, they continued and completed the procedure with two remaining arms, in the original configuration.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical inc.(isi), has not received the da vinci product with an alleged issue to perform failure analysis.The issue was confirmed during a site visit by an intuitive field services engineer (fse), requiring an adjustment or replacement of an essential system component.
|
| |
|