Catalog Number 0600520
Medical Device Problem Codes Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889); Expulsion (2933); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Skin Erosion (2075)
Date of Event 07/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2026, a patient underwent placement of a hickman/leonard/broviac central venous catheter in the left internal jugular vein for the administration of chemotherapy to treat leukemia.Following the procedure, on (b)(6) 2026, it was discovered that the broviac catheter had completely fractured.Prior to the fracture, the catheter had reportedly become twisted on multiple occasions due to patient movement in bed and had been subjected to tension while the patient was crawling; therefore, wear and tear associated with these factors was considered a possible contributing cause of the fracture.It was further reported that the catheter cuff was exposed outside the body, suggesting that it had become dislodged.The patient's current condition was unknown.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Broviac CV Catheter, Single-Lumen, 4.2F
Common Device Name
Chronic Catheters
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
Bard Shannon Limited -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50446950
Report Number3006260740-2026-06179
Device Sequence Number10376883
Product Code LJS
UDI-Device Identifier00801741051807
UDI-Public(01)00801741051807
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K830256
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number0600520
Device Lot NumberUnknown
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/21/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-