C.R. Bard, Inc. (BASD) -3006260740 Broviac CV Catheter, Single-Lumen, 4.2F; Chronic Catheters
|
Back to Search Results |
|
| Catalog Number |
0600520 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889); Expulsion (2933); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Skin Erosion (2075)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent placement of a hickman/leonard/broviac central venous catheter in the left internal jugular vein for the administration of chemotherapy to treat leukemia.Following the procedure, on (b)(6) 2026, it was discovered that the broviac catheter had completely fractured.Prior to the fracture, the catheter had reportedly become twisted on multiple occasions due to patient movement in bed and had been subjected to tension while the patient was crawling; therefore, wear and tear associated with these factors was considered a possible contributing cause of the fracture.It was further reported that the catheter cuff was exposed outside the body, suggesting that it had become dislodged.The patient's current condition was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|