H11: as the serial number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H11: manufacturing review: a dhr and manufacturing review were not required as the event was not classified as an unexpected event, the investigation did not identify any manufacturing related issues, and no repeat failures were found during the service history review.Investigation summary: the magnum driver serial number (b)(6) was received at the bip biomedizinische intstrumente und produkte gmbh.The magnum driver was visually inspected upon receipt, and it was found in normal condition.The cocking slide, cannula sled, stylet sled are worn out.The device was functionally tested and passed all the tests during evaluation.No other anomalies were identified.Therefore, the investigation is determined to be unconfirmed for the reported prematurely fires during cocking issue and the investigation is determined to be confirmed for the identified cocking slide, cannula sled, stylet sled are worn out issue.The root cause for reported prematurely fires during cocking issue cannot be determined as the problem could not be reproduced.The root cause for identified cocking slide, cannula sled, stylet sled are worn out issue could not be determined based on the provided information.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labelling, packaging, or use related issue.B3.The date received by manufacturer has been used for this field.G3, h6 (device, component, method, result, conclusion)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|