| Model Number |
SIF-Q180 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the small intestinal videoscope to the service center for a separate issue.During the device analysis, the service repair technician indicated that a foreign object was found in the nozzle and that the image was disturbed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: deformation of the electrical connector caused image disturbance when the connector was pushed.A cause for the foreign material found in the nozzle could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|