• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC Pump In Style Pro Traditional EN ES FR; Pump, Breast, Powered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC Pump In Style Pro Traditional EN ES FR; Pump, Breast, Powered Back to Search Results
Model Number 101047090
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 08/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2026, the customer alleged to medela llc that her pump in style pro was having trouble powering on.Additionally, she alleged that she developed mastitis and was prescribed antibiotics.
 
Additional Manufacturer Narrative
Customer service sent a replacement pump to the customer and asked for the return of her old pump for evaluation.In follow up with the complaint handler on (b)(6) 2026, the customer responded that she would return the pump for evaluation.Based on the results of our internal investigation (capa-2025-0027) including published literature, it cannot definitively be concluded that the pump caused or has contributed to the customer's reported mastitis.According to published clinical literature, mastitis incidence in lactating women, with or without a breast pump, can be as high as 33%.In fact, clinical guidelines recommend the use of a breast pump to facilitate the removal of breast milk during periods of mastitis.Additionally, complaint trends across all medela pumps show mastitis incidence consistently below 0.1%, far lower than the 3-20% occurrence in the general population of lactating women "[2]".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Pump In Style Pro Traditional EN ES FR
Common Device Name
Pump, Breast, Powered
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
alea raeber
1101 corporate drive
mchenry, IL 60050
+1(800)435-8316
MDR Report Key50449488
Report Number1419937-2026-00026
Device Sequence Number10310823
Product Code HGX
UDI-Device Identifier00020451470900
UDI-Public00020451470900
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number101047090
Device Catalogue Number101047090
Device Lot Number0000742013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/06/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/30/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-