SHIRAKAWA OLYMPUS CO., LTD. EVIS X1 VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-1500
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported no signal on serial digital interface (sdi) during inspection prior to use involving an olympus video system center.There was no patient injury reported.
 
Additional Manufacturer Narrative
The device was not returned to olympus for inspection, and the reported failure was confirmed.The customer contacted olympus technical assistance support (tac) by phone and reported the event.Technical support provided troubleshooting and was ultimately able to resolve the reported allegation.Based on the results of the investigation, the definitive root cause of the reported issue was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS X1 VIDEO SYSTEM CENTER
Common Device Name
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50450344
Report Number3002808148-2026-29843
Device Sequence Number9837962
Product Code FET
UDI-Device Identifier04953170414596
UDI-Public04953170414596
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K222861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCV-1500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/13/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-