ERIKA DE REYNOSA, S.A. DE C.V. 2008T HEMODIALYSIS SYSTEM WITHOUT CDX; Dialyzer, High Permeability With Or Without Sealed Dialysate System
|
Back to Search Results |
|
| Catalog Number |
190858 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
On (b)(6) 2026 a biomedical technician requested onsite service for machine verification following a hemodialysis (hd) patient incident on the 2008t hemodialysis machine.The patient coded and passed away on the machine.Attempts to obtain additional information were unsuccessful.The fresenius medical care field service technician (fst) evaluated the machine.The fst performed the ultrafiltration (uf) problem identification checklist according to p/n 507162 rev.G, performed uf pump calibration, verified 1.0cc/ml, functional checks were performed, and the machine was heat disinfected.The machine passed all functional testing.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.Clinical investigation: there is a temporal relationship between hd therapy utilizing the fresenius 2008t hemodialysis machine and the patient event of coding with subsequent death.However, there is no documentation to show a causal relationship between the adverse event and use of the 2008t machine.Additionally, there is no allegation of a machine malfunction or deficiency reported for this event.Although the cause of death was not obtained, ¿cardiac arrest accounts for a quarter of deaths among patients on dialysis.¿ there were no reported issues with the 2008t machine, and the machine passed all functional testing during evaluation.Based on the available information and no allegation or evidence of a malfunction or deficiency, the fresenius 2008t hemodialysis machine can be excluded as the cause of the patient¿s coding and death.
|
| |
|
Search Alerts/Recalls
|
|
|