SpectraWAVE, Inc. Starlight Imaging Catheter; Catheter, Intravascular, Plaque Morphology Evaluation
|
Back to Search Results |
|
| Model Number |
951 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
"it was reported that a starlight catheter was used for a coronary procedure.During use, red thrombus was seen inside the stent post pci in the lad, and treated by ballooning the thrombus to the wall and given integrilin bolus.The catheter had difficulty exiting the guide, and thought the catheter was going to kink.After several attempts, the angle of approach was changed, the catheter entered the vessel with trouble advancing into the pda, but was able to eventually cross with some maneuvering of the guidewire.This adverse event is being submitted because the patient experienced thrombus, requiring additional intervention.There was no alleged malfunction reported with the starlight catheter.".
|
| |
|
Event or Problem Description
|
|
"it was reported that a starlight catheter was used for a coronary procedure.During use, red thrombus was seen inside the stent post pci in the lad, and treated by ballooning the thrombus to the wall and given integrilin bolus.The catheter had difficulty exiting the guide, and thought the catheter was going to kink.After several attempts, the angle of approach was changed, the catheter entered the vessel with trouble advancing into the pda, but was able to eventually cross with some maneuvering of the guidewire.This adverse event is being submitted because the patient experienced thrombus, requiring additional intervention.There was no alleged malfunction reported with the starlight catheter.".
|
| |
|
Additional Manufacturer Narrative
|
|
Block a: patient information, not available from facility.Block c: not applicable for this device.Block d4: serial # not applicable for this device.Blocks d6 & d7: not applicable for this device.Block h3: the starlight catheter was not returned for evaluation; thus, no product investigation was performed.Block h6: device history record (dhr) was reviewed and all released units were manufactured to specification, with no ncrs or deviations contributing to the reported complaint.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.".
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental mdr 1: this report is being submitted as part of a retrospective review stemming from an fda form 483, and is captured in capa-52.Block a: patient information, not available from facility.Block c: not applicable for this device.Block d4: serial # not applicable for this device.Blocks d6 & d7: not applicable for this device.Block h3: the starlight catheter was not returned for evaluation; thus, no product investigation was performed.Block h6: device history record (dhr) was reviewed and all released units were manufactured to specification, with no ncrs or deviations contributing to the reported complaint.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.".
|
| |
|