|
The photo provided confirmed the report.The balloon of the device was observed to be ruptured.While the reported event was confirmed, the exact root cause of the reported issue could not be determined.Balloon rupture is an inherent risk of using this product
due to the nature of the device and the procedures it is being used for, the reported issue is a known complication of using the product.It is possible that procedural factors/anatomical features such as calcification, or plaque caused/contributed to the reported issue.The instructions for use (ifu) include the following warning regarding possible complications:
in common with other catheterization procedures, complications may occur.These may include but are not limited to:
infection, local hematomas, intimal disruption, arterial dissection, perforation and rupture, hemorrhage, arterial thrombosis, distal emboli of blood clots or arteriosclerotic plaque, air embolus, aneurysms, arterial spasms, arteriovenous fistula formation, balloon rupture or tip separation with fragmentation and distal embolization.Exercise caution when encountering extremely diseased vessels.Arterial rupture or balloon failure due to sharp calcified plaque may occur.The possibility of balloon rupture must be taken into account when considering the risks involved in the catheterization procedure.Due to the increased rate of occurrence of this issue, capa 2026-007 was created to document and investigate the failure.The device history record (dhr) is not available for review; therefore, a lot review could not be performed.
|