Brainlab SE Navigation SW Cranial EM (Version 1.1); Image Guided Surgery System/Instrument, Stereotaxic, Product Code: HAW
|
Back to Search Results |
|
| Model Number |
22240-06H |
| Medical Device Problem Codes |
Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A cranial surgery for a shunt placement in the brain, for drainage of cerebrospinal fluid (csf), has been performed with the aid of the brainlab navigation sw cranial em 1.1
after three navigated shunt placement attempts, without retrieving the expected csf, the surgeon determined that these had missed the ventricle and had deviated from the intended path.Navigation discontinued and surgeon placed the shunt successfully freehanded.According to the surgeon (treating clinician):
- the outcome of the surgery was successful as intended, without the aid of navigation.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement attempts.- there was no resulting harm or negative effect to the patient, also not due to the surgery/anesthesia prolongation of ca.30-60min.- there were further no remedial actions necessary, done or planned for this patient.There was also no prolongation of hospitalization.
|
| |
|
Additional Manufacturer Narrative
|
|
B2, h1:
a risk to the patient's health could not be excluded for these specific circumstances, since shunt placement attempts were performed in a different location in the brain than anticipated with the brainlab navigation involved, despite according to the surgeon (treating clinician):
- the outcome of the surgery was successful as intended, without the aid of navigation.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement attempts.- there was no resulting harm or negative effect to the patient, also not due to the surgery/anesthesia prolongation of ca.30-60min.- there were further no remedial actions necessary, done or planned for this patient.There was also no prolongation of hospitalization.A comprehensive investigation by brainlab regarding this specific event is currently ongoing and final conclusions are pending.Brainlab plans to issue a follow-up report upon completion of the investigation.H6,h7
a comprehensive investigation by brainlab regarding this specific event is currently ongoing and final conclusions are pending.Brainlab plans to issue a follow-up report to the fda upon completion of investigation.
|
| |
|
Search Alerts/Recalls
|
|
|