Model Number 22240-06H
Medical Device Problem Codes Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A cranial surgery for a shunt placement in the brain, for drainage of cerebrospinal fluid (csf), has been performed with the aid of the brainlab navigation sw cranial em 1.1 after three navigated shunt placement attempts, without retrieving the expected csf, the surgeon determined that these had missed the ventricle and had deviated from the intended path.Navigation discontinued and surgeon placed the shunt successfully freehanded.According to the surgeon (treating clinician): - the outcome of the surgery was successful as intended, without the aid of navigation.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement attempts.- there was no resulting harm or negative effect to the patient, also not due to the surgery/anesthesia prolongation of ca.30-60min.- there were further no remedial actions necessary, done or planned for this patient.There was also no prolongation of hospitalization.
 
Additional Manufacturer Narrative
B2, h1: a risk to the patient's health could not be excluded for these specific circumstances, since shunt placement attempts were performed in a different location in the brain than anticipated with the brainlab navigation involved, despite according to the surgeon (treating clinician): - the outcome of the surgery was successful as intended, without the aid of navigation.- there was no (direct or increased) risk of harm to a critical structure due to the deviating shunt placement attempts.- there was no resulting harm or negative effect to the patient, also not due to the surgery/anesthesia prolongation of ca.30-60min.- there were further no remedial actions necessary, done or planned for this patient.There was also no prolongation of hospitalization.A comprehensive investigation by brainlab regarding this specific event is currently ongoing and final conclusions are pending.Brainlab plans to issue a follow-up report upon completion of the investigation.H6,h7 a comprehensive investigation by brainlab regarding this specific event is currently ongoing and final conclusions are pending.Brainlab plans to issue a follow-up report to the fda upon completion of investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navigation SW Cranial EM (Version 1.1)
Common Device Name
Image Guided Surgery System/Instrument, Stereotaxic, Product Code: HAW
Manufacturer (Section D)
Brainlab SE
olof-palme-strasse 9
muenchen, 81829
GM  81829
Manufacturer (Section G)
Brainlab SE
olof-palme-strasse 9
muenchen, 81829
GM   81829
Manufacturer Contact
boutaina mediane
olof-palme-strasse 9
muenchen, 81829
GM   81829
MDR Report Key50455885
Report Number8043933-2026-00062
Device Sequence Number18759377
Product Code HAW
UDI-Device Identifier04056481132422
UDI-Public04056481132422
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K223288
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number22240-06H
Device Catalogue Number26553A
Device Lot NumberSW v 1.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2026
Supplement Date Received by Manufacturer07/28/2026
Initial Report FDA Received Date08/26/2026
Supplement Report FDA Received Date09/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-