Evaluation determined the product met specifications and no product nonconformance was identified.Visual: visual evaluation of the returned sample noted one opened (without original packaging), all silicone foley catheter attached to a drainage bag.Verified material number 119314 and manufacturing lot number ngkn0783.Visual inspection noted no obvious observations.Functional: inflated balloon with 10 ml of solution using the in house luer lock syringe and the catheter rested with no leaks noted.Deflated balloon passively with no cuffing or leaks noted.Flushed the drainage lumen with d.I.Water and it exited through the eyelets with no other leaks or defects noted.Flushed the sample port and filled the drainage bag with blue methyl solution and d.I, water and there were no leaks noted.Our team conducted a review of the reported event using all available information to ensure the continued safety and performance of the product.The investigation has been completed and determined that this issue represents a known failure mode, which has been previously assessed in the product risk management file in accordance with applicable regulatory standards and internal procedures.The associated risks continue to be actively monitored as part of our ongoing post market surveillance activities in compliance with regulatory requirements and company procedures.Based on the results of the investigation, no further action is required at this time.Corrections: d, h.Udi format updated with available product information per gudid.H11: sections a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|