Model Number 408719
Medical Device Problem Codes Failure to Capture (1081); High impedance (1291); Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
This lv lead had impedance of 3000 on all poles and no capture.Upon opening the pocket, the lead was found to be bent at a severe angle coming out of the lv port.The pocket was enlarged to allow the lead to lay in a straighter position and remains implanted.No adverse patient events were reported.Should additional information be received this file will be updated.
 
Additional Manufacturer Narrative
Combination product: yes.Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sentus ProMRI OTW QP L-85/49
Common Device Name
LV lead
Manufacturer (Section D)
Biotronik SE & Co. KG
woermannkehre 1
berlin 12359
GM  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
+1(877)245-9800
MDR Report Key50456373
Report Number1028232-2026-03965
Device Sequence Number10310581
Product Code OJX
UDI-Device Identifier04035479148959
UDI-Public(01)04035479148959(17)260531
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P070008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model Number408719
Device Catalogue Number408719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2026
Supplement Date Received by Manufacturer08/31/2026
Initial Report FDA Received Date08/26/2026
Supplement Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-