| Model Number |
BL11145 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported an influx of pressure.Due to high pressure when going from phaco to i/a resulted in an influx of fluid to the eye.The procedure was halted with workaround, but patient had to return to finish the surgery.No harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The fluidics and compressor modules were replaced by the field service engineer on site.The module serial numbers (b)(6) and (b)(6) were evaluated with no evidence of condition that could or would lead to this complaint.The (b)(6) plungers worked as expected.The (b)(6) had no evidence of over pressurization, all compressor functional values were in specification.The field service engineer provided no evidence of reproducing complaint.The device history was reviewed and found to meet manufacturing specification.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
|
| |
|