A peritoneal dialysis (pd) patient contact reported to fresenius medical care (fme) technical service (ts) on (b)(6) 2026 that the liberty cycler alarmed with an m31 air detected in cassette warning in fill 5 of 5.The patient (pt) was connected during incident and fluid was seen.Upon follow-up conducted on (b)(6) 2026 the patient contact stated that on (b)(6) 2026 the patient¿s liberty cycler started to alarm with m31 air detected in cassette warning in fill 5 of 5.The patient contact stated that after disconnecting the patient from the cycler with the assistance of the fresenius technical support the patient noticed that there was some fluid on the floor.The contact confirmed that the patient uses and that they were using a 7.25% baxter solution bag (lot unknown) for the last fill.The contact stated that the fluid leaked from the connection of the luer lock connector to the cassette.Per patient contact there were no external leaks from the fresenius solution bag or any connections prior, during, or after the event.The patient contact also confirmed that there was no fluid intrusion inside the cycler as the result of the reported event.The contact stated that they did not know what caused the leak and confirmed that there were no known delivery issues or damage to the delivery boxes the solutions and supplies were delivered in.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient was not able to complete treatment on the day of the reported event.The patient was able to resume treatment on the cycler with no further issues the following day.The sample is not available to be returned to the manufacturer for physical evaluation.
|