(b)(4).Date sent: 8/26/2026.D4: batch#: 876d61.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pve35a device was returned with no apparent damage and with a reload loaded on the device.The reload was received unfired.The device opened as expected.The manual override door out of position and missing; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequence firings.The bailout system was reset and then an attempt to fire the device was made, however, the device would not fire.Upon evaluation, the pcb board was noted to be non-functional.No conclusion could be reached as to what may have caused the pcb board to be non-functional.The event described could not be confirmed as the reload returned unfired and the device opened and closed as expected.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished pve35a, with lot number: a9995j, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number: 876d61, and no non-conformances were identified.
|