(b)(4).Date sent: 8/26/2026
d4: batch (b)(4).Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ec45a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.During functional testing, the device exhibited difficulty opening and the firing trigger show slow return, the firing trigger was slightly stuck and needed manually slightly push to fully return the firing trigger to it's home position.The device was disassembled, and the spring pawl extension was noted to be out of its intended position, interfering with the trigger plate subassembly not allowing to open the trigger as expected.The damage to the spring pawl extension is potentially related to a manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.A manufacturing record evaluation was performed for the finished device batch number (b)(4), and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|