Model Number LDA210Q/58
Medical Device Problem Codes Over-Sensing (1438); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
During a remote follow-up, decreased sensing and far-field p-wave oversensing (ffpwos) were observed on the right ventricular (rv) lead.An x-ray was performed and the rv lead was found to be dislodged.No intervention has been performed but a revision is anticipated.The patient is stable with no consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Common Device Name
Permanent Defibrillator Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50459527
Report Number2017865-2026-17952
Device Sequence Number10264203
Product Code NVY
UDI-Device Identifier05414734507325
UDI-Public(01)05414734507325(10)A000188528(17)290531
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLDA210Q/58
Device Lot NumberA000188528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1