On july 29, 2026, axora representative was informed of a post-procedure adverse event involving a hydro thermal ablation (hta) case on (b)(6) 2026.The procedure was completed successfully with no reported fluid loss; however, thermal burns to the patient's labia and vaginal side were identified two weeks later when the patient called reporting pain.The patient required follow-up medical treatment for wound care.
|
|
Axora received additional information on 8/12/26 from the physician during a discussion with axora medical's physician.The treating physician stated that the hta procedure was on (b)(6) 2026, and the thermal injury was noticed 2 weeks later when the patient complained of pain at the introitus.At the most recent visit, healing has progressed but is not complete.The treating physician admitted she did not continuously view the cervix during the procedure and thus was not able to observe fluid loss during the procedure.She recalled the fluid loss indicator fluctuated frequently, though it never reached the 10 ml level, which would have triggered the alarm and shut down.Additionally, the physician reported using a weighted speculum, placed in the posterior vaginal rather than graves or pederson speculum, both of which are better at keeping the lateral and anterior walls of the vaginal apart without the use of additional retractors.Based on the available information, there was a possibility of intermittent fluid loss during the procedure, never exceeding 10 ml.An actual fluid loss was not noticed by the treating physician because the cervix and vagina were not monitored continuously during the procedure.The use of a weighted speculum may have made it more difficult to view the cervix during the procedure.As a result, small amounts of heated fluid may have leaked unnoticed without triggering the safety alarm.Per the physician, a company representative observed her next case and confirmed the equipment operated properly with no issues.The ifu instructs the user to "during ablation, the saline temperature continues to rise until 90°c is reached, and the user must continuously:
monitor the cervical seal for fluid loss and to confirm a tight cervical seal.Maintain a stable procedure sheath position throughout the procedure.Monitor the screen for fluid loss level."
warnings and cautions related to an event of this nature are included in the ifu for genesys hta, including but not limited to:
the physician must maintain control of the procedure sheath (i.E., not hand off to another individual) for the duration of the treatment to avoid compromising the cervical seal.A compromise of the cervical seal could result in fluid leakage through the cervix, which could result in thermal injury to surrounding tissue.Per the labeling review, the fluid loss alarm signals a loss of at least 10ml of fluid.Fluid losses in excess of 10 ml may occur in cases when the alarm is triggered.If a fluid loss occurred for the third time after the temperature reached 50°c.For safety reasons, the system will automatically shut down.The user should expect a shutdown if the controller alerts a 3rd fluid loss, as the controller performed as intended, alerting the user of a potential leak and to check for possible uterine perforation.As with all endometrial ablation procedures, serious injury or death can occur.Thermal injury to adjacent tissue is explicitly listed as a potential adverse event in the section other adverse events.
|