| Catalog Number |
14-533021 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using the zyston curve inserter, the surgeon positioned the cage between the vertebral bodies in the straight configuration and attempted to rotate the dial on the handle to release the cage.However, the adjacent silver-colored component rotated together with the dial, preventing the dial from loosening.The dial was eventually loosened with the assistance of pliers.A boot-shaped impactor was used to rotate the mechanism, and the procedure was successfully completed.There was no patient or procedural impact.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d4: lot & udi numbers, d9, and h3.Additional information in h4 and h6: investigation type, findings, and conclusions.A review of the dhr was conducted and found no indications of manufacturing issues.This event likely cannot be attributed to manufacturing or design related issues.An inspection of the part shows the device has disassembled.No damage to the separated components was observed.The complaint is confirmed for zyston inserter for the failure of disassembly.The failure could possibly be attributed to damage over time and repeated usage of the instruments leading to the disassembly.This device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report 3012447612-2026-00229.
|
| |
|