Catalog Number 3505004BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Seroma (2069); Impaired Healing (2378); Fluid Discharge (2686)
Date of Event 08/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent a primary breast augmentation procedure with a mentor memorygel breast implant gel breast prosthesis developed a wound on her left breast post implantation.The wound had a pimple-like appearance.She subsequently developed a full thickness wound with exposed implant.Additionally, imaging indicated left breast prosthesis rupture.The patient also developed baker grade iii capsular contracture in her left breast.As a result, the patient underwent explantation of the left breast prosthesis.There was a small amount of serous appearing fluid in the breast pocket.A sample was sent to pathology.The implant was not replaced as the patient needs time to heal before another surgery.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left breast wound with exposed implant, breast prosthesis rupture, capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MemoryGel Breast Implant
Common Device Name
Prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
Mentor Texas
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
Mentor Texas
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
+1(949)788-2078
MDR Report Key50459818
Report Number1645337-2026-09555
Device Sequence Number15945807
Product Code FTR
UDI-Device Identifier00081317000532
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2013
Device Catalogue Number3505004BC
Device Lot Number5842423
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date08/26/2026
Supplement Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
Patient RaceWhite
-
-