| Catalog Number |
ECP0110GV |
| Medical Device Problem Codes |
Break (1069); Failure to Cycle (1142); Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6)2026, a patient underwent a vacuum-assisted breast biopsy procedure using an encor probe.During the procedure, the needle failed to close correctly after use.It was reported that the device could have cut the tissue; however, the samples obtained were particularly small despite the dense breast tissue.It was further reported that the distal end of the needle was permanently damaged and was noticed at the time of device removal from the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr captured an incorrect manufacturing location (g1).The correct manufacturing location for this device is ausa medical devices private limited.No new event information is being reported in this supplement.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.G1, g3.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|