• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Davol Inc., Sub. C.R. Bard, Inc. -1213643 Phasix Mesh; Surgical Mesh

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Davol Inc., Sub. C.R. Bard, Inc. -1213643 Phasix Mesh; Surgical Mesh Back to Search Results
Catalog Number 1190500
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Unspecified Infection (1930); Scar Tissue (2060)
Date of Event 07/24/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Per information provided: (b)(6) 2012 - patient was diagnosed with recurrent ventral hernia, incarcerated incisional hernia, thereby underwent open repair with explant of synthetic mesh, implant of ventrio st mesh (device 1).Per op notes, "through a midline incision, previously placed synthetic mesh was identified and was dissected.Then a midline fascia was identified and repaired by placing a ventrio st mesh (device 1) into the defect and was sutured." (b)(6) 2013 - patient was diagnosed with recurrent incisional hernia, thereby underwent open repair with implant of ventrio st mesh (device 2).Per op notes, "hernia defect in the abdominal region was identified and was dissected.A ventrio st mesh (device 2) was placed and secured in place with sutures to the previous mesh (device 1) in the midline and was tacked to cover the entire defect." (b)(6) 2015 - patient was diagnosed with recurrent ventral hernia, thereby underwent open repair.Per op notes, "purulent drainage was noted from the middle portion of the midline incision site." (b)(6) 2016 - patient was diagnosed with abdominal pain, thereby underwent repair with implant of biologic mesh.Extensive intra-abdominal small bowel adhesive disease from prior ventral hernia mesh repair, with a quite large piece of mesh.Small bowel densely adherent to this mesh in the rlq of the abdomen and suspected to be the source of abdominal pain.(b)(6) 2017 - patient was diagnosed with chronic abdominal pain, inflammation around the intra-abdominal mesh, thereby underwent explant of ventrio st mesh (device 1 and 2), implant of phasix flat mesh (device 3).Per op notes, "scar tissue in the midline was identified.Dense adhesions were noted.The ventrio st mesh (device 1 and 2) was found turned around in few places and had adhesions which was then lysed.The ventrio st mesh (device 1 and 2) were excised.The midline fascia was then closed by placing phasix flat mesh (device 3)." (b)(6) 2017 - patient was diagnosed with chronic abdominal wall wound, thereby underwent open repair with explant of phasix flat mesh (device 3).Per op notes, "the scar tissue was identified and excised.Some portions of phasix flat mesh (device 3) might be reabsorbable.The exposed portion of the infected phasix flat mesh (device 3) was excised.".
 
Additional Manufacturer Narrative
Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-aug-2016) is considered to be a best estimate.This emdr represents the phasix flat mesh (device 3).Additional supplemental emdrs were submitted to represent the ventrio st mesh (device 1 & 2).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Phasix Mesh
Common Device Name
Surgical Mesh
Manufacturer (Section D)
Davol Inc., Sub. C.R. Bard, Inc. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
Bard Shannon Limited -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
+1(800)556-6756
MDR Report Key50459855
Report Number1213643-2026-01114
Device Sequence Number13277700
Product Code OOD
UDI-Device Identifier00801741077708
UDI-Public(01)00801741077708(17)180828(10)HUAV0561
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K142818
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2018
Device Catalogue Number1190500
Device Lot NumberHUAV0561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date08/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/15/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-