TERUMO BCT, INC. Spectra Optia®; Separator, Automated, Blood Cell And Plasma, Therapeutic
|
Back to Search Results |
|
| Model Number |
61000 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Treatment started at 1300.At 1335 fluid leak detector alarmed indicating fluid in centrifuge area.Treatment paused, centrifuge door opened and noted blood leak.Treatment was terminated with ebl "[estimated blood loss]" 185ml.All component of kit were double checked prior to treatment.Upon inspection all components were properly loaded into the machine.Breakage was noted in tubing distal to hexagon lock.Manufacturer was notified of kit malfunction, and information sent per their request.Optia removed from service for deep cleaning and inspection by htm "[healthcare technology management]"/vendor.
|
| |
|