B3: date of event is estimated.The inogen team attempted to contact the patient for further information three times with no response.The unit was returned with a system hot complaint, which was confirmed during testing.The unit "system hot "shutdown because fan was misaligned and contacting the accumulator likely due to high-impact incident, possibly from the unit being dropped.Also, stuck feed waste valve caused by debris inside it and the muffler being filled with dust, which resulted in a blockage in the feed port.Furthermore, the psa cycle (according to the data log) being high (14) is an indication the zeolite is getting contaminated by moisture.Uip, top housing & lower housing were also replaced due to cosmetic damage.
|