• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nobel Biocare AB NobelParallel CC RP 4.3x10mm; Endosseous Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nobel Biocare AB NobelParallel CC RP 4.3x10mm; Endosseous Dental Implant Back to Search Results
Catalog Number 37972
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Failure to osseointegrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NobelParallel CC RP 4.3x10mm
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Nobel Biocare AB
dimbovägen 2
karlskoga 69151
SW  69151
MDR Report Key50463792
Report Number2027971-2026-121131
Device Sequence Number15996381
Product Code DZE
UDI-Device Identifier07332747091519
UDI-Public(01)07332747091519(10)12126238(17)231222
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/26/2026
Report Date (Section F) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/22/2023
Device Catalogue Number37972
Device Lot Number12126238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/21/2026
Event Location Outpatient Treatment Facility
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/26/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-