• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Dexcom, Inc. Stelo Glucose Biosensor System; Continuous Glucose Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Dexcom, Inc. Stelo Glucose Biosensor System; Continuous Glucose Monitor Back to Search Results
Model Number 9500-174
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Stelo Glucose Biosensor System
Common Device Name
Continuous Glucose Monitor
Manufacturer (Section D)
Dexcom, Inc.
6340 sequence dr.
san diego CA 92121
Manufacturer (Section G)
Dexcom, Inc.
pmt 818
persiaran cassia selatan 3
bandar cassia 14110
MY   14110
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
+1(858)200-0200
MDR Report Key50465122
Report Number3004753838-2026-236862
Device Sequence Number21306703
Product Code SAF
UDI-Device Identifier00386270004345
UDI-Public(01)00386270004345(11)251001(17)270331(10)1825286010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K234070
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9500-174
Device Catalogue NumberSTP-XN-001
Device Lot Number1825286010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/11/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age40 YR
Patient SexFemale
-
-