| Model Number |
5816 |
| Medical Device Problem Code |
Unexpected Mode Switch (4076)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented for a follow-up in clinic.Interrogation revealed that the pacemaker had entered backup vvi mode.The patient was then admitted to the hospital and attempts to restore the pacemaker to normal operation through reprogramming were unsuccessful due to normal battery depletion.No intervention or changes were performed.No patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported implant date is an estimate.The implant date was reported as follow, ¿the pacemaker was implanted in 2018, but the exact date is unknown¿.
|
| |
|