| Model Number |
103606-620 |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Senseonics was made aware of an incident where user reported an event where the cgm showed results that are different from blood glucometer measurements.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The user reported discrepancies between bg and sg readings.Escalation analysis indicated that bg values met sg trends well considering the expected lag between the sg and bg inherent to sensing technology.Customer care intended to recommend that the user continue using the system and to enter calibrations when glucose trends were not changing rapidly; however, further follow-up with the user was not possible, as the user remained unresponsive to multiple communication attempts.As per the data management system (dms) review, the system remained in active use with up-to-date information.
|
| |
|