U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Senseonics Inc. Eversense Sensor; Eversense Continuous Glucose Monitor System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Senseonics Inc. Eversense Sensor; Eversense Continuous Glucose Monitor System Back to Search Results
Model Number 103606-620
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
Senseonics was made aware of an incident where user reported an event where the cgm showed results that are different from blood glucometer measurements.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
The user reported discrepancies between bg and sg readings.Escalation analysis indicated that bg values met sg trends well considering the expected lag between the sg and bg inherent to sensing technology.Customer care intended to recommend that the user continue using the system and to enter calibrations when glucose trends were not changing rapidly; however, further follow-up with the user was not possible, as the user remained unresponsive to multiple communication attempts.As per the data management system (dms) review, the system remained in active use with up-to-date information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Eversense Sensor
Common Device Name
Eversense Continuous Glucose Monitor System
Manufacturer (Section D)
Senseonics Inc.
20451 seneca meadows parkway
germantown MD 20876 -700
Manufacturer (Section G)
Senseonics Inc.
20451 seneca meadows parkway
germantown MD 20876 -700
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876--700
MDR Report Key50465813
Report Number3009862700-2026-02711
Device Sequence Number9898264
Product Code SBA
UDI-Device Identifier00817491024305
UDI-Public00817491024305
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number103606-620
Device Catalogue NumberFG-7502-00-301
Device Lot Number01G117S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/19/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexUnknown
-
-