• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION Synclara Cough System; DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION Synclara Cough System; DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT Back to Search Results
Model Number M08575
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
Baxter received a report that synclara system, non-ft sw,hc,na had cracked filter.This occurred at home before use.There was no patient injury or death reported with this event.
 
Additional Manufacturer Narrative
The baxter technical support found the patient circuit kit needed to be replaced.The synclara cough system is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness.The synclara cough system provides a noninvasive therapy designed for use by patients, caregivers, and healthcare providers.The system will simulate a cough to remove secretions in patients with a compromised peak cough flow.Possible adverse conditions listed in the device instructions for use include swallowing too much air resulting in the likelihood of vomiting and aspiration.The use of a face mask may result in discomfort, vomiting, or breathlessness in some patients.Close supervision throughout the treatment is necessary when this product is used by children or patients with physical limitations or impaired cognitive abilities.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this product.It is unknown if the customer performs preventative maintenance on their products.A replacement patient circuit kit with filter was sent to the customer.Based on this information, no further actions are required at this time.Although there was no reported injury with this event, if the report of a cracked filter were to recur, it could potentially cause serious injury or death.Therefore baxter is reporting this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Synclara Cough System
Common Device Name
DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM SINGAPORE
1 yishun avenue 7
singapore, north east NA 76892 3
SN   768923
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
+1(224)270-5355
MDR Report Key50465978
Report Number3008145987-2026-00070
Device Sequence Number10063083
Product Code NHJ
UDI-Device Identifier00887761984554
UDI-Public(01)00887761984554
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K192143
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Type of Report Initial
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberM08575
Device Catalogue NumberM08575
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-