H3: the tubing set, two clamps, one clip, the device, and an open pouch were returned in a double plastic bag.Upon receipt, the pouch was open on three of its four sides.No traces of contamination were observed on the returned components.No damage was noted to the tubing set, outer tube, outer hub, or inner hub.However, the inner shaft between the inner and rotating hub was slightly bent.The inner assembly rotated by hand without binding.The device was loaded into a handpiece and oscillated up to 5,000 rpm.During blade oscillation, wobbling was observed.The device failed due to the defective condition upon return and the inability to rotate properly.The complaint was confirmed, due to an out of specification condition.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|