|
Mdr / initial 1 of 2.Unomedical hereby submits this supplemental report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates (b)(4) (ic retrospective complaint review) and qp-ic-(b)(4) (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.Additional information - this electronic medical device report (emdr) is being submitted to include the below:
h6: investigation results under type of investigation, investigation findings, investigation conclusions
h11: investigation summary
complaint investigation results:
electronic quality management system (eqms) search:
a query was run in the electronic quality management system (eqms) on (b)(6)2026 against lot number 5389794 and similar malfunction codes leakage from connection between the tubing connector and the infusion set (cannula part/base piece),tubing connector is cracked, split, deformed, leakage may occur, leak between tubing and site connector ¿ trace moisture / minor wetness.The review confirmed that lot 5389794 and the identified failure mode are not associated with any nonconforming reports (ncr)s or corrective and preventive action (capa)s of the same or similar nature.The complaint can be closed referencing the investigation (as it addresses the malfunction).Similar complaints search:
a query was run in the electronic quality management system (eqms) on (b)(6)2026 against lot number criteria equal 5389794 and similar malfunction codes: leakage from connection between the tubing connector and the infusion set (cannula part/base piece), tubing connector is cracked, split, deformed, leakage may occur, leak between tubing and site connector ¿ trace moisture / minor wetness.The count of complaints is 1.The complaint number is: complaint (b)(4).Visual evidence review:
no photo was provided to support visual confirmation of the reported issue.Conclusion:
unable to perform visual verification; assessment based on available documentation only.Corrective and preventive action (capa) determination assessment:
based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts)
complaint unconfirmed:
following investigation, the reported failure could not be confirmed for this complaint.The complaint was unconfirmed based on analysis.Based on the available information, this event is deemed to be a reportable malfunction
to date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number
reporting site: 8021545
manufacturing site: 3003442380.
|