C.R. Bard, Inc. (BASD) -3006260740 Hickman CV Catheter, Dual-Lumen, 7F; Chronic Catheters
|
Back to Search Results |
|
| Catalog Number |
0600570 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260); Material Separation (1562); Stretched (1601); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a catheter placement procedure using a hickman/leonard/broviac central venous catheter.After a period of use, on (b)(6) 2026, inadequate securement and line management, along with continuous stretching, twisting, and kinking of the line, resulted in an almost complete tear/rupture of the white lumen of the dual lumen tunneled line.It was further reported that the issue was identified when blood leakage was observed on the bed.The catheter was subsequently repaired using a repair kit.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.However, the return of the sample is pending.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|