A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 11-aug-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the system biometric identifier was not working.A technical support specialist (tss) verified the biometric login issue, restarted the lumidigm services, rebooted the station, and confirmed sensor errors persisted despite configuration checks.The field service engineer (fse) reviewed settings, reseated and tested the biometric reader, and replaced the reader; however, the issue remained.After collaborating with medbank support, corrective actions were performed, and biometric login functionality was restored.Multiple successful login tests were completed by the customer and additional users, confirming normal operation and resolution of the issue.Login issue, restarted the lumidigm services, rebooted the station, and confirmed sensor errors persisted despite configuration checks.The system functioned as intended after the field service engineer repaired the device.
|