| Model Number |
500001004001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ pro medstation¿ es, user synchronized a username with another user badge.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 09-apr-2026 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the user synchronized a username with another user's badge.A technical support specialist (tss) reviewed the badge authentication records and determined that the badge scan was associated with the correct user account.The tss concluded that an incorrect fingerprint scan may have caused the issue and advised re-enrolling the fingerprint by intentionally failing fingerprint authentication multiple times to access password login, then updating the biometric identification settings through user preferences and confirmed that the issue was resolved and confirmed to close the case.The system functioned as intended after the technical support specialist investigated the device.
|
| |
|
Search Alerts/Recalls
|