• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1250 SYNCHRONY 2
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code Head Injury (1879)
Date of Event 07/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
Hearing performance with the device has been affected after the user sustained a trauma to the head on the opposite side.
 
Event or Problem Description
Hearing performance with the device has been affected after the user sustained a trauma to the head on the opposite side.Reimplantation is planned, but no date has been scheduled yet.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Additional information: based on the information received, a mechanical damage of the stimulator electronics, which is typical for a severe external impact to the housing, appears likely.However, to determine and confirm an exact root cause of failure cause a device investigation of the explanted device is necessary.Re-implantation is planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL Elektromedizinische Geraete GmbH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key50470608
Report Number9710014-2026-00796
Device Sequence Number10264534
Product Code MCM
UDI-Device Identifier09008737369026
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/19/2025
Device Model NumberMi1250 SYNCHRONY 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Supplement Date Received by Manufacturer08/24/2026
Initial Report FDA Received Date08/27/2026
Supplement Report FDA Received Date09/23/2026
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age5 YR
Patient SexMale
-
-