| Model Number |
Mi1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
07/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hearing performance with the device has been affected after the user sustained a trauma to the head on the opposite side.
|
| |
|
Event or Problem Description
|
|
Hearing performance with the device has been affected after the user sustained a trauma to the head on the opposite side.Reimplantation is planned, but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the information received, a mechanical damage of the stimulator electronics, which is typical for a severe external impact to the housing, appears likely.However, to determine and confirm an exact root cause of failure cause a device investigation of the explanted device is necessary.Re-implantation is planned.
|
| |
|