Alcon Research, LLC - Huntington AcrySof Multipiece IOL; Intraocular Lens Back to Search Results
Model Number MA60AC
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative reported with the description of torn or broken haptic.Additional information received and stated that during phaco with intraocular lens (iol) implant procedure haptic was broken.There was patient contact but no patient harm.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.The product was not returned.A batch record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the medical device file were met.The account indicated the use of a qualified associated cartridge and viscoelastic.The handpiece information was not provided.It is unknown if a qualified handpiece was used.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.It is unknown if a qualified handpiece was used.The instructions for use (ifu) instructs: company foldable intraocular lens are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ophthalmic viscosurgical device (ovd)) combination.The instructions for use (ifu) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The instructions for use (ifu) instructs to completely fill the cartridge with ophthalmic viscosurgical device (ovd) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ophthalmic viscosurgical device (ovd) in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ophthalmic viscosurgical device (ovd), which may result in damage.The instructions for use (ifu) instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ophthalmic viscosurgical device (ovd)-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.Lenses with a 6.5 millimetre optic diameter may need to be pre-folded for easy entry into the back of the cartridge.Lens may be pre-folded by gently applying pressure to the edge of the lens while holding the central optic with forceps the trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The investigation has been completed based on current information.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaints are reviewed and monitored for adverse trends on a monthly basis.No adverse trends have been observed associated with the reported product and event.No further action has been identified for this reported event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was no returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the medical device file were met.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AcrySof Multipiece IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
Alcon Research, LLC - Huntington
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50470801
Report Number1119421-2026-01765
Device Sequence Number18685002
Product Code HQL
UDI-Device Identifier00380652552622
UDI-Public(01)00380652552622(17)270912(11)220914(10)15480782
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.250
Device Lot Number15480782
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/27/2026
Date Device Manufactured09/14/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DuoVisc VISCOELASTIC SYSTEM; MONARCH II IOL Delivery System, Cartridge B
Patient Age73 YR
Patient SexUnknown
-
-