| Model Number |
C315HIS02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the implant procedure, catheter hemostatic valve was not working properly.Catheter was not used and was replaced but the same issue was observed with the second catheter.Second catheter was not used and was replaced.The catheters were returned to the manufacturer, analysed, and tested out of specification.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the delivery catheter was returned and analyzed.The septum detached from the hub of the delivery catheter.The septum adhesive released of the hub of the delivery catheter.The septum of the catheter was damaged.Visual analysis of the delivery catheter indicated damage during use.The analyst noted the delivery catheter was returned without the dilator.The septum had detached and was pushed down into the hub of the delivery catheter.There was adhesive present within the hub of the delivery catheter.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|