| Model Number |
AG606972A |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Necrosis (1971)
|
| Date of Event |
09/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Incision ports on abdomen and flanks not healing 10 months post procedure that combined liposuction and bodytite.According to the patient, after he complained to the doctor that the incisions were not closing, he was treated with antibiotics and drainage attempt was performed.8 months post tx he received steroids injections into each incision port.The patient also reported contour irregularity near his navel which is attributed to an excessive liposuction, not related to the bodytite device.
|
| |
|
Additional Manufacturer Narrative
|
|
No technical issues were revealed upon system inspection.Although the doctor did not provide photos/ccq, the case was investigated based on the photos provided by the patient.It appears the doctor persistently went back too close to the incision ports while pulsing, which led initially to an extensive thermal damage at the incision ports.This in turn led to the observed necrosis which prevents the healing.The root cause was established to be an incorrect technique, not congruent with the ifu.Per latest update, the patient turned to another clinic for excision revision.
|
| |
|