| Catalog Number |
F30000162 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
A user facility reported a noise that occurred during support for 28 days.A grinding-like noise was observed when the speed was increased above 7100 rpm, making it impossible to increase the rpm further.The noise was normal at speeds of 7000 rpm or below.The first kit was initially used at 6600 rpm.As the patient¿s condition deteriorated, attempts were made to increase the rotational speed to 7500 rpm; however, this was not possible due to the abnormal noise.After the kit was replaced, the second kit reportedly produced the same abnormal noise whenever the speed was increased above 7100 rpm from the beginning of use.The user requested a determination of whether this noise is normal or abnormal, and an analysis of the cause of the noise.An unspecified patient injury and death were reported.The complaint sample was not available for product investigation.
|
| |
|
Event or Problem Description
|
|
A user facility reported a noise that occurred during support for 28 days for a patient on extracorporeal membrane oxygenation support due to pulmonary failure.A grinding-like noise was observed when the speed was increased above 7100 rpm, making it impossible to increase the rpm further.The noise was normal at speeds of 7000 rpm or below.The first kit was initially used at 6600 rpm.As the patient¿s condition deteriorated, attempts were made to increase the rotational speed to 7500 rpm; however, this was not possible due to the abnormal noise.After the kit was replaced, the second kit reportedly produced the same abnormal noise whenever the speed was increased above 7100 rpm from the beginning of use.The user requested a determination of whether this noise is normal or abnormal, and an analysis of the cause of the noise.There was no indication of patient serious injury.The patient expired on (b)(6) 2026 due to an unreported cause.The complaint sample was not available for product investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical review: multiple attempts were made to acquire further details concerning this patient¿s expiration; however, no additional information was obtained.The cause of death was not reported, and no further information has been provided.It is unknown if a temporal relationship exists between ecmo support utilizing the novalung console and the xlung kit 230 and the patient¿s death.In the absence of the required information, the cause of the patient¿s death cannot be determined.It is well established critically ill patients on ecmo support carry a high risk of mortality due to underlying pathology and/or the patient¿s response to ecmo support (1).It is within reason to believe that the cause of the patient¿s death was multifactorial and more than likely not related to abnormal console pump noise.Regardless, the cause of this patient¿s death cannot exclude ecmo support or the performance of any fme/xenios product(s) or device(s).Based on the available information, there was no specific allegation or objective evidence of any fresenius product(s) or device(s) deficiency or malfunction that caused or contributed to this patient¿s adverse event.Should additional information become available related to any fresenius device(s) and/or product(s), a clinical investigation will be reassessed accordingly.Additional information: a3, a4, b2, b5 the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|